M600L Multi-parameter Analyzer [EC-T Version]
Features
GENERAL FEATURES
- Color High contrast LCD touchscreen, 7.0 inches.
- Temperature Unit Options: Choose between ºC or °F.
- Reset feature automatically resumes all settings back to factory default options.
- IP54 waterproof.
Professional-Moldable, Multi-channel
- Modular four Channel: Four measurements channels can be configured in any combination of the modules.
- Standard configuration: 1 EC-T module. Support the expansion of other modules, such as pH module, pX module, EC-T module, EC-F module, DO-F module, DO-P module.
- The multi-channel allows measurements of 1, 2, 3, or 4 parameters simultaneously, either inthe same or in different samples.
- According to the module, the parameters can be pH, mV, EC, Resistivity, TDS, Sal., ISE, pX, DO, DO Saturation and Temp.
Smart operation, Auto-recognition, High throughput
- Intelligent operating system, with user management, method management, method management, sample management, electrode management, data management, GMP management and other functions.
- User three-level authority management and password management.
- Method management function, built-in 16 methods and custom measurement methods.
- Multi-reading feature allows auto-read, timed-read and continuous-read.
- Calibration reminder and forced calibration function.
- Verification reminder and forced verification function.
- Calibration and measurement alarm limit setting.
- Connection with automatic sampler to realize the automatic measurement and batch detection of samples.
Data management, Audit trial
- Data Storage 1000 sets of each parameter (GLP-compliant).
- Support for GMP management, the GMP logs achieve the tracking and analysis of the the data (compliant with FDA 21 CFR Part 11 ).
- RS-232 communication for print and the print format can be alternate.
- USB communication for PC software.
- USB drive for method export and import.
- Support for USB drive for data export with CSV or PDF format.
- USB Barcode Scanner Support.
EC-T module
- It shows conductivity, Resistivity, TDS, Salinity and Ash.
- 1-5 points calibration with GB and International standard recognition.
- Temperature compensation type (none, linear, pure water).
- Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.
- Data storage of 1000 sets of EC/Resistivity/TDS/Salinity/Ash measurement results, 100 sets of calibration data, and 1000 sets of EC standard solution verification data.
Includes
- DJS-1VTC Conductivity Electrode(K=0.6, ATC)
- 1413μS/cm EC standard solution, 50mL, 1 vial
- REX-6 electrode stand
- Touch pen
Specifications
Model | M600L Multi-parameter Analyzer [EC-T Version] | |
Module | EC-T module | |
Parameters | pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation) | |
Conductivity | Range | 0.000μS/cm~3000mS/cm |
| Resolution | 0.001μS/cm minimum, changed with range |
| Accuracy | ±0.5%FS |
Resistivity | Range | 5.00Ω·cm~100.0MΩ·cm |
| Resolution | 0.01Ω·cm minimum, changed with range |
| Accuracy | ±0.5%FS |
TDS | Range | 0.000 ppm~1000ppt |
| Resolution | 0.001ppm minimum, changed with range |
| Accuracy | ±0.5%FS |
Salinity | Range | (0.0~80.0) ppt, (0.00~8.00) %NaCl, (2.000~42.000)psu |
| Resolution | 0.1ppt, 0.1%NaCl, 0.001psu |
| Accuracy | ±1ppt |
Temperature | Range | (-10.0~135.0)ºC, (14.0~275.0)ºF |
| Resolution | 0.1ºC, 0.1ºF |
| Accuracy | ±0.3ºC (0.0~60.0)ºC;±1.0 ºC (Other Range) |
Environment | Temperature: (0~40) ºC Relative Humidity: ≤85% | |
Power | AC Adapter(Input:100-240V AC; Output: 20V DC) | |
Dimensions | 240×310×98mm | |
Weight | About 2.0kg | |